The Goldilocks zone for 3T MRS studies using semi-LASER: Determining the optimal balance between repetition time and scan time using FSL-MRS
Le résumé fourni par la source
Motivation: A short TR is often used in clinical MRS brain studies, leading to T1-weighting of metabolites and inaccurate T1 correction. Goal(s): To determine the optimal balance between scan time and TR that minimizes T1-weighting effects when using semi-LASER MRS. Approach: SNR and metabolite concentrations were compared for TRs of 2, 5 and 8s using FSL-MRS. Results: A TR of 5s provides a good balance of scan time, SNR and signal recovery. For a similar scan time, TR of 2s leads to incomplete signal recovery and thus heavy T1 weighting, while a TR of 8s results in complete signal recovery but reduced SNR. Impact: For MRS studies using semi-LASER, our work informs scientists and clinicians on the issues of using a short TR, and recommends the optimal scan parameters to use while implementing the newly available FSL-MRS analysis package.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- The Goldilocks zone for 3T MRS studies using semi-LASER: Determining the optimal balance between repetition time and scan time using FSL-MRS
- Date Crossref
- 26/11/2024
- Éditeur
- ISMRM
- Type
- proceedings-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.