Making myelin water imaging “normal”: Robust acquisition for cervical cord with low peripheral nerve stimulation
Le résumé fourni par la source
The standard myelin water imaging (MWI) sequence for cervical spinal cord, 3D gradient and spin echo (GRASE), was modified to allow scanning in “normal mode” (limiting imaging parameters that may cause physiologic stress) for patients who may not tolerate peripheral nerve stimulation or tissue heating due to MR-conditional implants or medical issues. Traditional 32-echo GRASE was replaced with 48-echo GRASE in normal mode. Myelin water fraction maps from the new sequence had better repeatability than the standard sequence. 48-echo GRASE is the recommended sequence for MWI in spinal cord, particularly for subjects who cannot be scanned outside of normal mode.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Making myelin water imaging “normal”: Robust acquisition for cervical cord with low peripheral nerve stimulation
- Date Crossref
- 03/08/2023
- Éditeur
- ISMRM
- Type
- proceedings-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.