Development of electrochemiluminescence immunochromatographic test strips for dengue virus

0Citations signalées — pas une note de qualité
6Institutions déclarées
3Pays d’affiliation déclarés

Résumé fourni par la source

Introduction Dengue virus (DENV) infection poses a serious threat to global public health. Rapid and sensitive point-of-care diagnostic methods are urgently needed for early detection and epidemic control. Methods The dengue virus monoclonal antibody NS1-McAb1# was labeled as the labeled antibody and sprayed onto the binding pad using ruthenium tripyridine as the signal probe, and the dengue virus monoclonal antibody DN-REAB-G2-006 was fixed on the detection pad as the capture antibody. To mitigate potential interference from other substances in the sample, the sample pad and conjugate pad was pretreated with a solution containing heterophilic blocking reagent, and Elecsys buffer was selected as the optimal co-reactant. After dropping the sample onto the sample pad, chromatography and immunoreaction were performed sequentially. Then, cyclic voltammetry scanning was carried out in the test zone at a rate of 0.1 V/s for electrochemiluminescence (ECL) detection. A novel detection method was established by optimizing various parameters. Result The ruthenium-labeled antibody was prepared by reacting Ru-NHS ester with NS1-McAb1# at a molar ratio of 50:1 for 10 h. Elecsys Buffer was selected as the co-reactant, with a sample-to-buffer volume ratio of 2:1. Whatman AE 100 membrane was used as the detection pad with a length set to 15 mm. The concentration ratio of labeled antibody to capture antibody was 1:2. The chromatographic time after sample loading was 10 min. Under these optimized conditions, the detection method achieved a minimum detection limit of 20.76 pg/mL, exhibited good specificity, with an intra-assay repeatability coefficient of variation of 4%, a storage stability coefficient of variation of 0.83%, and an inter-assay stability coefficient of variation of 0.81%. Clinical sample validation showed a Kappa value of 0.625 compared with nucleic acid testing results. Discussion The developed ECLIA test strip provides a portable, rapid, and sensitive detection method for DENV, with potential applications in on-site testing across multiple scenarios. Although preliminary clinical evaluation showed acceptable concordance with nucleic acid testing, further large-scale validation using diverse clinical specimens is necessary to fully establish its diagnostic utility in real-world settings.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Development of electrochemiluminescence immunochromatographic test strips for dengue virus
Date Crossref
27/08/2026
Éditeur
Frontiers Media SA
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Mosquito-borne diseases and controlAdvanced biosensing and bioanalysis techniquesBiosensors and Analytical Detection

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.