Fat and fat‐free mass measurement agreement by dual‐energy X‐ray absorptiometry versus bioelectrical impedance analysis: Effects of posture and waist circumference

11Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Background: Bioelectrical impedance analysis (BIA) operates under the assumption that the conductor has a uniform cylindrical shape. However, this assumption may be violated if measures are taken in the seated position, especially in people with a high waist circumference (WC). Aims: The aims of this research were to determine whether posture (supine, standing, and seated) and WC affect agreement between BIA and dual-energy X-ray absorptiometry (DXA) measures of fat mass (FM) and fat-free mass (FFM). Materials & Methods: ). Body composition was measured using BIA in the supine, standing, and seated positions and by DXA while supine. Intraclass correlation coefficient (ICC) analyses with two-way mixed effects and absolute agreement were performed to determine agreement. Results: Point estimates were excellent for FM and FFM while supine, excellent for FM and good for FFM while standing, and moderate for FM and good for FFM while seated. BIA measures in the supine position resulted in the narrowest 95% confidence intervals compared with other positions. Better agreement was observed across all positions in participants with a WC below the median (118.3 cm). Discussion: Despite the potential pragmatic value of measuring with BIA in a seated position, the results of this analysis demonstrate the poorest agreement between DXA and BIA methods, especially in individuals with high WC. Conclusion: Findings from this study demonstrate that BIA, particularly when measured in a supine position, can serve as a viable alternative to DXA for measuring body composition in people with obesity.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Fat and fat‐free mass measurement agreement by dual‐energy X‐ray absorptiometry versus bioelectrical impedance analysis: Effects of posture and waist circumference
Date Crossref
11/03/2024
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Body Composition Measurement TechniquesNutrition and Health in AgingElectrical and Bioimpedance Tomography

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.