Comparison of EPID portal dosimetry verification and RadCalc dose verification for VMAT treatment plans

1Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : pl. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Introduction. Dosimetry verification is required before starting each treatment. The legal regulations do not clearly define one method of plan verification. Therefore, it is allowed to perform measurements (electronic portal imaging device [EPID]) or calculations using an independent system. Portal dosimetry using EPID matrices was compared with the RadCalcTM system v. 7.1.4.1, performing independent dose distribution calculations. Materials and methods. Treatment plans were made for 150 patients treated with the photon 6MV VMAT technique. Three groups of patients were studied: those treated for breast cancer, those treated for prostate cancer, and those irradiated to the prostate area with nodes. Then, the dosimetry verification was carried out on the accelerator using the EPID portal and compared with the independent RadCalc software calculation results. Results. Comparison of tumor proportion score (TPS) vs. EPID and vs. RC calculations for breast, prostate, and prostate with nodes showed no significant statistical differences. Regardless of the size (volume) of the clinical target volume (CTV) area, no significant difference was observed, although there was a greater agreement for large CTVs compared to small ones. Similarly, there was no significant difference in the compared methods based on depth, but there was a better agreement for small depths than large ones. Conclusions. Verification methods in the study groups showed compliance of the measured (EPID) and calculated (RadCalc) doses with the values planned in the TPS. This confirms that verification for patients treated with radiotherapy can be performed with any of these methods. However, for radiosurgical techniques, it is better to use the EPID method because the RadCalc method may give false negative results.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Comparison of EPID portal dosimetry verification and RadCalc dose verification for VMAT treatment plans
Date Crossref
04/01/2024
Éditeur
VM Media Group sp. z o.o
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Advanced Radiotherapy TechniquesRadiation Effects and DosimetryRadiation Therapy and Dosimetry

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.