Are all ventilators for NIV performing the same? A bench analysis

7Citations signalées, ce qui n’est pas une note de qualité
4Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : es. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Global pandemic due to COVID-19 has increased the interest for ventilators´ use worldwide. New devices have been developed and older ones have undergone a renewed interest, but we lack robust evidence about performance of each ventilator to match appropriate device to a given patient and care environment. The aim of this bench study was to investigate the performance of six devices for noninvasive ventilation, and to compare them in terms of volume delivered, trigger response, pressurization capacity and synchronization in volume assisted controlled and pressure support ventilation. All ventilators were tested under thirty-six experimental conditions by using the lung model ASL5000® (IngMar Medical, Pittsburgh, PA). Two leak levels, two muscle inspiratory efforts and three mechanical patterns were combined for simulation. Trigger function was assessed by measurement of trigger-delay time. Pressurization capacity was evaluated as area under the pressure-time curve over the first 500 ms after inspiratory effort onset. Synchronization was evaluated by the asynchrony index and by incidence and type of asynchronies in each condition. All ventilators showed a good performance, even if pressurization capacity was worse than expected. Leak level did not affect their function. Differences were found during low muscle effort and obstructive pattern. In general, Philips Trilogy Evo/EV300 and Hamilton C3 showed the best results. NIV devices successfully compensate air leaks but still underperform with low muscle effort and obstructive lungs. Clinicians´ must have a clear understanding of the goals of NIV both for devices´ choice and set main parameters to achieve therapy success.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Are all ventilators for NIV performing the same? A bench analysis
Date Crossref
31/07/2023
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Respiratory Support and MechanismsIntensive Care Unit Cognitive DisordersChronic Obstructive Pulmonary Disease (COPD) Research

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.