The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system
Rattachement africain : au. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Purpose: This study assessed the delivered dose accuracy in pancreas SBRT by incorporating the real-time target position determined using an in-house position monitoring system. Methods and materials: An online image-based position monitoring system, SeedTracker, was developed to monitor radiopaque marker positions using monoscopic x-ray images, available from the Elekta XVI imaging system. This system was applied to patients receiving SBRT for pancreatic cancer on the MASTERPLAN Pilot trial (ACTRN 12617001642370). All patients were implanted pre-treatment with at least three peri-tumoral radiopaque markers for target localisation. During treatment delivery, marker positions were compared to expected positions delineated from the planning CT. The position tolerance of ±3mm from the expected position of the markers was set to trigger a gating event (GE) during treatment. The dosimetric impact of position deviations and actual dose delivered with position corrections was assessed by convolving the plan control point dose matrices with temporal target positions determined during treatment. Results: Eight patients were treated within this study. At least one GE was observed in 38% of the treatment fractions and more than one GE was observed in 10% of the fractions. The position deviations resulted in the mean(range) difference of -0.1(-1.1 - 0.4)Gy in minimum dose to tumour and 1.9(-0.1- 4.6)Gy increase to Dmax to duodenum compared to planned dose. In actual treatment delivery with the patient realignment, the mean difference of tumour min dose and duodenal Dmax was reduced to 0.1(-1.0 - 1.1)Gy and 1.1 (-0.7 - 3.3)Gy respectively compared to the planned dose. Conclusions: The in-house real-time position monitoring system improved the treatment accuracy of pancreatic SBRT in a general-purpose linac and enabled assessment of delivered dose by incorporating the temporal target position during delivery. The intrafraction motion impacts the dose to tumour even if target position is maintained within a 3mm position tolerance.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system
- Date Crossref
- 28/11/2022
- Éditeur
- Frontiers Media SA
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Où se fait cette recherche
-
UNSW Sydney pays non établi dans la noticeUniversité ou école supérieure
-
Ingham Institute Department of Medical Physics pays non établi dans la noticeOrganisation à but non lucratif
-
The University of Sydney pays non établi dans la noticeUniversité ou école supérieure
-
Cancer Australia pays non établi dans la noticeOrganisme public
-
University of New South Wales South Western Clinical School pays non établi dans la noticeUniversité ou école supérieure
-
School of Physics Institute of Medical Physics pays non établi dans la noticeUniversité ou école supérieure
-
Department of Radiation Oncology pays non établi dans la noticeInstitution
UNSW Sydney, Department of Medical Physics — Ingham Institute et The University of Sydney, avec 4 autres affiliations.
Une affiliation ne permet pas de déduire la nationalité d’un auteur.