Aller au contenu principal
2021 article

End‐diastolic segmentation of intravascular ultrasound images enables more reproducible volumetric analysis of atheroma burden

7Citations signalées — pas une note de qualité
9Institutions déclarées
4Pays d’affiliation déclarés

Résumé fourni par la source

BACKGROUND: Volumetric intravascular ultrasound (IVUS) analysis is currently performed at a fixed frame interval, neglecting the cyclic changes in vessel dimensions occurring during the cardiac cycle that can affect the reproducibility of the results. Analysis of end-diastolic (ED) IVUS frames has been proposed to overcome this limitation. However, at present, there is lack of data to support its superiority over conventional IVUS. OBJECTIVES: The present study aims to compare the reproducibility of IVUS volumetric analysis performed at a fixed frame interval and at the ED frames, identified retrospectively using a novel deep-learning methodology. METHODS: IVUS data acquired from 97 vessels were included in the present study; each vessel was segmented at 1 mm interval (conventional approach) and at ED frame twice by an expert analyst. Reproducibility was tested for the following metrics; normalized lumen, vessel and total atheroma volume (TAV), and percent atheroma volume (PAV). RESULTS: ; p < 0.001) and PAV (-0.12 ± 0.59% vs. -0.34 ± 1.34%; p < 0.001). Results were similar when the analysis focused on the 10 mm most diseased segment. The superiority of the ED approach was due to a more reproducible detection of the segment of interest and to the fact that it was not susceptible to the longitudinal motion of the IVUS probe and the cyclic changes in vessel dimensions during the cardiac cycle. CONCLUSIONS: ED IVUS segmentation enables more reproducible volumetric analysis and quantification of TAV and PAV that are established end points in longitudinal studies assessing the efficacy of novel pharmacotherapies. Therefore, it should be preferred over conventional IVUS analysis as its higher reproducibility is expected to have an impact on the sample size calculation for the primary end point.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
End‐diastolic segmentation of intravascular ultrasound images enables more reproducible volumetric analysis of atheroma burden
Date Crossref
17/08/2021
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Coronary Interventions and DiagnosticsCardiac Imaging and DiagnosticsCerebrovascular and Carotid Artery Diseases

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.