Acute angle closure after implantable contact lens insertion unresponsive to surgical peripheral iridectomy
Rattachement africain : nz. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
We present a case of acute angle closure that occurred after insertion of an implantable contact lens (ICL). The apparent papillary-block angle closure did not resolve after 2 patent iridotomies and a surgical iridectomy, but did respond to pupil dilation (not constriction). Ultrasound biomicroscopy revealed abnormally large and irregular ciliary processes that may have contributed to the unusual behavior of the ICL-iris complex. The condition resolved after the ICL was replaced by one with a smaller haptic diameter. Routine ultrasound biomicroscopic assessment of the ciliary body anatomy preoperatively and ICL haptic positioning postoperatively may identify risk factors that could predispose ICL patients to acute angle closure.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Acute angle closure after implantable contact lens insertion unresponsive to surgical peripheral iridectomy
- Date Crossref
- 01/04/2008
- Éditeur
- Ovid Technologies (Wolters Kluwer Health)
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.