CMOS readout system for a double-sided germanium strip detector
Rattachement africain : us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
A wide variety of applications require a hard X-ray or gamma ray detector which combine both good spatial resolution and energy resolution. Double-sided solid-state strip detectors provide this capability, We report on the development of CMOS electronics designed for photon detection with strip detectors. These electronics include a low noise preamplifier, semi-gaussian shaping amplifier, discriminator, and peak detection circuitry. All circuits are designed to operate at low power. Circuits have been duplicated in both NMOS and PMOS to provide both polarities of signals. We have constructed an 8/spl times/8 channel system to test these prototype chips. Power consumption for the preamplifier through peak-detector circuit is 4 mW/channel. The test system has a conversion gain of /spl sim/35 mV/fC, system noise (equivalent noise charge) of ENC100:1 in both NMOS and PMOS circuits.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- CMOS readout system for a double-sided germanium strip detector
- Date Crossref
- 22/11/2002
- Éditeur
- IEEE
- Type
- proceedings-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.