Improved Visualization of Middle Ear Prostheses using Photon-Counting Detector CT: A Cadaver Study
Rattachement africain : us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Abstract Background Evaluation of middle ear prostheses requires high-resolution imaging to assess position and structural integrity. Purpose To assess the diagnostic performance of photon-counting-detector (PCD)-CT for middle ear prosthesis imaging in cadaveric specimens. Materials and Methods Nine different middle ear prostheses were implanted in five fresh-frozen whole-head specimens, both in correct anatomic position and displaced to mimic malposition. Specimens were scanned on PCD-CT (NAEOTOM Alpha, Siemens) and energy-integrating-detector (EID)-CT (SOMATOM Force, Siemens), reconstructed at 0.2mm and 0.4mm, respectively. Eighteen cases (9 prostheses, 2 positions) per scanner were evaluated by two neuroradiologists (36 readings), blinded to scanner and position, who scored prosthesis position, reader confidence, and overall image quality. Agreement with the reference standard and between readers was compared. Results PCD-CT showed higher diagnostic confidence compared with EID-CT, while accuracy and inter-reader agreement did not demonstrate significant differences between scanners. Correct readings, PCD-CT versus EID-CT: lateral contact, 31/36 (86.1%; 95% CI: 58.3%-94.4%) vs. 29/36 (80.6%; 55.6%-91.7%), p = 0.500; medial contact, 30/36 (83.3%; 55.6%-94.4%) vs. 26/36 (72.2%; 50.0%-83.3%), p = 0.250. PCD-CT identified all cases of complete contact loss (score 3), while EID-CT missed one reading per contact. Inter-reader agreement may be higher with PCD-CT vs EID-CT (lateral: κw: 0.804 [0.578-0.941] vs. 0.571 [0.302-0.778], p = 0.250; medial: 0.684 [0.191-0.934] vs. 0.481 [0.024-0.835], p = 0.250). Reader confidence was significantly higher with PCD-CT; mean paired differences for Readers 1 and 2 were 9.4 (1.8-17.1; Holm-adjusted p = 0.019) and 23.1 points (13.4-32.7; p < 0.001) for lateral contact, and 16.4 (10.0-22.8; p < 0.001) and 17.8 points (6.8-28.8; p = 0.007) for medial contact. Readers preferred PCD-CT image quality in 16/18 (88.9%; 65.3%-98.6%) and 17/18 cases (94.4%; 72.7%-99.9%), respectively. Conclusion PCD-CT was associated with significantly higher reader confidence and preferred image quality, without significant differences in diagnostic accuracy or inter-reader agreement. These findings suggest PCD-CT is a promising modality for postoperative middle ear evaluations that warrants further investigations.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Improved Visualization of Middle Ear Prostheses using Photon-Counting Detector CT: A Cadaver Study
- Date Crossref
- 15/09/2026
- Éditeur
- Oxford University Press (OUP)
- Type
- journal-article