Fast Patient-Specific Breast CT Dosimetry: 22-Fold Acceleration of Monte Carlo MGD Estimation

0Citations signalées — pas une note de qualité
0Institutions déclarées
0Pays d’affiliation déclarés

Résumé fourni par la source

Accurate patient-specific mean glandular dose (MGD) estimation in breast computed tomography (BCT) requires anatomically realistic models, but high-resolution patient-derived phantoms impose high computational demands on Monte Carlo (MC) dosimetry. We previously developed and validated an EGSnrc-based framework for synchrotron propagation-based phase-contrast BCT (PB-PCT) for the first-in-human study at the Australian Synchrotron. With participant imaging commencing, this study aimed to optimize phantom spatial resolution and MC efficiency while maintaining MGD accuracy, for a rapid patient-specific dosimetry. Four patient-specific heterogeneous breast phantoms with varying volume and glandularity were generated from PB-PCT images of mastectomy specimens acquired at 35 keV. The highest-resolution phantom for each specimen served as the reference. Phantoms were downsampled by increasing in-plane voxel size and slice thickness, and MGD was recalculated using EGSnrc/DOSXYZnrc. A 4.5% difference from reference MGD was adopted as the accuracy criterion. Primary photon histories were then reduced to determine the minimum required for acceptable MC precision. A common voxel size of 0.30 x 0.30 x 3.00 mm3 maintained MGD within 4.5% of reference for all four phantoms while reducing voxel number by approximately 20-31-fold. MGD was more sensitive to in-plane voxel size than slice thickness, supporting anisotropic downsampling. Reduced-resolution phantoms enabled a 15-fold reduction in photon histories to 2 x 10^8 while maintaining acceptable statistical uncertainty. Combined optimization reduced computation time from approximately 1600 to 72 min, a 22-fold acceleration. Patient-specific phantoms can be substantially downsampled while maintaining MGD accuracy and markedly reducing computational burden. This optimization provides a foundation for future real-time patient-specific BCT dosimetry.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

La source scientifique ouverte est momentanément indisponible.

Sujets associés

Digital Radiography and Breast ImagingRadiation Dose and ImagingMedical Imaging Techniques and Applications

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.