A118 APAL on CN-6000 complements verifynow for multi-timepoint P2Y12 monitoring in neuroendovascular care

0Citations signalées, ce qui n’est pas une note de qualité
4Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : jp. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Background/Introduction The need for antiplatelet therapy is increasing in neuroendovascular treatments. To assess the risk of perioperative complications, we need reliable monitoring of P2Y12 inhibition. VerifyNow is common, but single-use cartridges can make repeated testing expensive. The Automated Blood Coagulation Analyzer CN-6000 allows platelet aggregation testing (APAL) in the routine coagulation lab (with PT, APTT, D-dimer). In the CN-6000 setting, the measurement cost for APAL was around tenfold lower than for a PRU test (Uematsu K et al. Ann Clin Lab Sci. 2020), making serial testing realistic. Objective/Purpose We studied whether APAL can support VerifyNow in daily practice. Methods/Case Description We reviewed 124 patients who underwent elective neuroendovascular procedure while on antiplatelet therapy. Blood was taken at four points: the day before the procedure, days 1–3 after the procedure, around 1 month, and at bleeding or thrombotic events. We compared four tests: APAL and 10 µM ADP maximum aggregation (MA) on CN-6000, PRU and % inhibition on VerifyNow. We assessed time-course changes, test correlations, and a practical APAL–PRU reference. Results/Findings PRU, % inhibition, and APAL changed significantly over time. 10 µM ADP MA changed little. Correlation with PRU was higher for APAL (r = 0.55) than for 10 µM ADP MA (r = 0.42). A practical reference showed APAL ≈ 8.3 corresponds to PRU = 240, a commonly used high on-treatment platelet reactivity threshold. Discussion/Conclusions/Learning Points For centers that use VerifyNow, measuring APAL on CN-6000 can: clarify borderline or unexpected PRU results and serve as a backup when PRU is invalid (instrument/sample error); make serial testing at key timepoints (pre-operation, days 1–3, ~1 month, event-driven) low-cost and practical; use PRU for pre-operation, with APAL available at every timepoint.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
A118 APAL on CN-6000 complements verifynow for multi-timepoint P2Y12 monitoring in neuroendovascular care
Date Crossref
01/09/2026
Éditeur
BMJ Publishing Group Ltd.
Type
proceedings-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.