Identifying optimal dose head CT acquisition techniques through cluster analysis from 904,209 scans

0Citations signalées — pas une note de qualité
8Institutions déclarées
3Pays d’affiliation déclarés

Résumé fourni par la source

Abstract Objectives To identify clinically established and frequently utilized routine head computed tomography (CT) acquisition techniques in children and adults that achieve lower radiation. Materials and methods The level of the analysis is at the CT scan (series) level. CT acquisition parameters were analyzed from a large, international CT dose registry. Scans were analyzed separately in adults and children, and by approach (helical or axial), resulting in four strata. K-means clustering, an unsupervised machine learning approach, assigned the scans into clusters based on the following acquisition parameters: Tube current, voltage, collimation, scan length, and for helical scans, pitch. Effective patient diameter, unadjusted and patient size-adjusted dose-length product (DLP), and volume CT dose index (CTDI vol ) were compared across clusters. Results 864,182 CT scans in adults and 40,027 scans in children between January 1, 2015, and March 11, 2021, were included in the analysis. Scans were classified into seven helical and six axial clusters in adults, six helical and four axial clusters in children. In all four strata, the mean size-adjusted DLP varied more than two-fold, and up to 3.4-fold, across the clusters. For example, for adult helical scans (the largest strata), the size-adjusted DLP ranged from 372 mGy·cm to 998 mGy·cm (relative dose 2.7) across clusters. Clusters with the lowest radiation generally utilized lower tube current and voltage, but protocols utilized various approaches to result in lower doses. Conclusion There remains a large variation in the radiation dose within frequently used head CT acquisition techniques. Clustering analysis enables identifying protocols with lower doses. Key Points Question Is there a large variation in the doses used for routine head CT, and can we identify best-practice, lowest dose approaches for scanning? Findings Across each stratum (adult/pediatric, helical/axial), the mean size-adjusted DLP varied by more than two-fold across the CT scan clusters, reflecting significant differences in radiation . Clinical relevance Many routine head CT acquisition techniques use excessive radiation doses. Standardizing practice to the lowest dose clusters would result in significant reductions in patient dose.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Identifying optimal dose head CT acquisition techniques through cluster analysis from 904,209 scans
Date Crossref
02/09/2026
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Radiation Dose and ImagingAdvanced Radiotherapy TechniquesAdvanced X-ray and CT Imaging

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.