Leadless Atrial Pacing at Bachmann's Bundle Guided by Electroanatomic Mapping and Intracardiac Ultrasound
Résumé fourni par la source
INTRODUCTION: Leadless pacemakers (LPs) offer advantages over transvenous systems, but atrial LPs currently target the right atrial appendage (RAA), risking atrial dyssynchrony. METHODS AND RESULTS: We report five patients in whom electroanatomic mapping and intracardiac ultrasound guided LP implantation into Bachmann's bundle (BB), a site shown to improve interatrial conduction and reduce atrial fibrillation risk. In four of five patients, BB fixation was successful with low pacing thresholds and short P-wave duration. One patient required conversion to RAA pacing due to high BB pacing thresholds. No intraprocedural complications occurred. CONCLUSIONS: This series demonstrates the feasibility of leadless BB pacing, though technical challenges warrant further study.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Leadless Atrial Pacing at Bachmann's Bundle Guided by Electroanatomic Mapping and Intracardiac Ultrasound
- Date Crossref
- 30/08/2026
- Éditeur
- Wiley
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.
Institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.