ACPSEM guideline: recommendations for a fluoroscopic system quality assurance program

0Citations signalées, ce qui n’est pas une note de qualité
10Institutions déclarées
3Pays d’affiliation déclarés

Rattachement africain : au, ie, nz. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Fluoroscopic systems are increasingly used for complex image-guided procedures across medical disciplines in Australia and New Zealand. The associated radiation exposure risks to patients and staff necessitate robust quality assurance (QA) programs. However, no national guidance has previously existed for a standardised fluoroscopy QA program in the Australia and New Zealand context. To provide evidence-informed recommendations for the quality assurance of fluoroscopic systems in Australia and New Zealand, addressing the absence of national standards and supporting consistent, high-quality clinical practice. This guideline was developed by the ACPSEM Fluoroscopy QA Working Group, established in February 2019 by the ACPSEM Radiology Specialty Group. Recommendations were informed by an expert-driven review of international standards and protocols, including those from the AAPM, IAEA, EFOMP, and IEC, adapted for Australia and New Zealand practice through working-group discussions and reviews. The draft guideline underwent stakeholder consultation, including circulation to ACPSEM members and relevant professional organisations. The guideline outlines recommended QA tests for medical physicists at acceptance, commissioning, and periodic intervals, along with routine quality control procedures for radiographers. Key areas include radiation dose output assessment, automatic dose rate control verification, image quality evaluation, personal protective equipment integrity testing, and clinical dose audits. A quality review meeting framework is also provided. This guideline establishes a nationally consistent framework for fluoroscopy QA that complements local regulatory requirements. Its adoption will support patient safety, staff radiation protection, and optimal equipment performance across healthcare facilities in Australia and New Zealand.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
ACPSEM guideline: recommendations for a fluoroscopic system quality assurance program
Date Crossref
27/08/2026
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Radiation Dose and ImagingRadiology practices and educationDigital Radiography and Breast Imaging

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.