Impact of x-ray CT calibration steps on: DVC and P-DVC displacement uncertainties

0Citations signalées — pas une note de qualité
3Institutions déclarées
2Pays d’affiliation déclarés

Résumé fourni par la source

Abstract X-ray tomography can be used to measure displacement fields from in situ or ex situ mechanical tests. The present work aims to quantify displacement uncertainties resulting from different reconstruction steps implemented in computed tomography (CT). Two samples made of polyamide and aluminum alloy were imaged in situ by acquiring radiograph sets corresponding to two tomographic turns. From these data sets, the standard displacement uncertainties were assessed via digital volume correlation (DVC) and projection-based DVC. Careful calibration of the CT geometry, accounting for beam hardening and correcting dark-and flat-field variations, was crucial to reach sub-decivoxel uncertainties.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Impact of x-ray CT calibration steps on: DVC and P-DVC displacement uncertainties
Date Crossref
01/09/2026
Éditeur
IOP Publishing
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Advanced X-ray and CT ImagingMedical Imaging Techniques and ApplicationsOrthopaedic implants and arthroplasty

BNTIC News n’est pas le producteur de ces données. Exploration à la demande auprès d’OpenAlex, avec contrôle bibliographique public par Crossref. Aucun service payant requis, aucune réponse conservée. Sources et limites.