Concept and simulation of a MEMS cycloidal mass analyzer
Résumé fourni par la source
This study investigates the miniaturization pathway of cycloidal mass spectrometry, a technique distinguished by its inherent perfect-focusing properties, and introduces a planar, stacked-layer MEMS-based cycloidal mass analyzer. A comprehensive numerical analysis was conducted to elucidate the effects of key design parameters, namely electric sector geometry, electrode geometry, and operating vacuum, on ion focusing and mass separation performance. The results indicate that appropriate optimization of geometric dimensions and electrode arrangement density effectively mitigates electric field distortion, thereby improving ion beam focusing and resolution. Configured with an electric sector size of 32.60 mm × 41.10 mm × 6.85 mm and 83 electrode pairs, and operated under a vacuum of 1 × 10 −3 Pa, the device achieves full width at half maximum (FWHM) values of approximately 0.03 Da for light ions (H 2 + , He + ) and about 0.40 Da for heavier ions (Ar + , CO 2 + ). Across the m / z range of 2–50 Da, the analyzer maintains a resolution greater than 99 and an ion detection efficiency exceeding 92%, demonstrating robust separation and transmission performance over a wide mass-to-charge range.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Concept and simulation of a MEMS cycloidal mass analyzer
- Date Crossref
- 14/08/2026
- Éditeur
- Springer Science and Business Media LLC
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.
Institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.