A high centrifugal force-enhanced Ziehl–Neelsen method for improved detection of Mycobacterium tuberculosis

0Citations signalées — pas une note de qualité
4Institutions déclarées
2Pays d’affiliation déclarés

Résumé fourni par la source

Ziehl-Neelsen (ZN) smear microscopy remains central to tuberculosis (TB) diagnosis and treatment monitoring; however, its sensitivity is limited by incomplete recovery of Mycobacterium tuberculosis during pre-analytical processing. This study evaluated whether modifying centrifugation force and duration improves bacillary recovery and ZN smear microscopy performance. Laboratory experiments were conducted using M. tuberculosis H37Rv suspensions and clinical sputum specimens. Following NALC-NaOH treatment, samples were centrifuged at 2,000, 3,000, and 6,000 × g for 40 min. The effect of centrifugation duration was assessed at 3,000 × g by comparing 20 and 40 min using the same M. tuberculosis H37Rv cultures and the same clinical sputum specimens at both time points, ensuring paired measurements within each sample type. Smear positivity and ZN smear grading were evaluated from replicate smears and analyzed using non-parametric statistical tests, with significance set at p < 0.05. In M. tuberculosis H37Rv suspensions, no significant differences in smear positivity or grading were observed across centrifugal forces (p = 0.368 and p = 0.212, respectively). In clinical sputum specimens, smear positivity did not differ significantly across forces (p = 0.716), whereas ZN smear grading increased significantly with higher centrifugal force (p = 0.0051). At 3,000 × g, extending centrifugation time from 20 to 40 min did not significantly affect smear positivity in either sample type (both p = 1.000). In contrast, ZN smear grading increased from 1+ to 2+ in clinical specimens with extended centrifugation time (p = 0.016), while no change was observed in M. tuberculosis H37Rv suspensions. These findings indicate that increasing centrifugal force may enhance bacillary concentration in clinical sputum, resulting in improved smear grading without a corresponding increase in detection rate. Extending centrifugation time has limited impact on smear positivity. Optimization of pre-analytical centrifugation parameters may improve ZN smear microscopy performance in clinical specimens.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
A high centrifugal force-enhanced Ziehl–Neelsen method for improved detection of Mycobacterium tuberculosis
Date Crossref
02/07/2026
Éditeur
Public Library of Science (PLoS)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Image Processing Techniques and ApplicationsTuberculosis Research and EpidemiologyBacterial Identification and Susceptibility Testing

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.