Axial length correction in retinal thickness measurements in diabetic retinopathy

0Citations signalées — pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

Abstract Purpose To use the RS-1 Glauvas (NIDEK) system to evaluate the effect of axial length (AL) correction on retinal thickness (RT) with optical coherence tomography (OCT). Study design Observational, cross-sectional study. Methods Between November 2023 and April 2024, 165 eyes from patients with diabetes underwent OCT imaging. RT was measured with and without AL-based magnification correction with the Early Treatment Diabetic Retinopathy Study (ETDRS) grid, which comprises three concentric circles (diameters; 1, 3, and 6 mm), with the inner and outer rings divided into superior, temporal, nasal, and inferior sectors. Measurements were calculated using two settings: a fixed AL of 24 mm (uncorrected) and the individually measured AL for each eye (corrected). The difference between uncorrected and corrected RT measurements was analyzed using Bland–Altman analysis. Linear regression analysis was performed to assess the association between AL and the measurement difference. Results The mean AL was 23.9 ± 1.2 mm. Bland–Altman analysis for central subfield thickness (CST) showed a small mean difference between corrected and uncorrected measurements (bias −0.34 µm; 95% limits of agreement −6.32 to 5.63 µm). Linear regression demonstrated a significant positive association between AL and the difference in CST measurements (R 2 = 0.70, p < 0.0001), indicating that uncorrected measurements tended to underestimate and overestimate RT in eyes with shorter and longer AL, respectively. Conclusion AL differences affect the effective OCT scan area and may influence RT measurements. Applying AL correction minimizes measurement bias caused by transverse magnification and improves the accuracy of RT assessment.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Axial length correction in retinal thickness measurements in diabetic retinopathy
Date Crossref
29/06/2026
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Retinal Diseases and TreatmentsGlaucoma and retinal disordersRetinal and Macular Surgery

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.