International collaborative study for the calibration of a proposed World Health Organization International Standard for thromboplastin, human, recombinant: a report from the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)—Scientific and Standardization Committee (SSC)/ISTH working group on prothrombin time/international normalized ratio standardization

0Citations signalées — pas une note de qualité
8Institutions déclarées
6Pays d’affiliation déclarés

Résumé fourni par la source

Background: The availability of International Reference Preparations for thromboplastins, later on referred to as International Standards (ISs) for thromboplastins is essential for calibration of prothrombin time (PT) measurement systems and for the international normalized ratio (INR). It is proposed to replace the current fifth IS for thromboplastin, human, recombinant (coded rTF/16) and the fifth IS for thromboplastin, rabbit, plain (coded RBT/16) by a single IS for defining a reference measurement system for PT/INR tests, in agreement with the concept of metrological traceability. Objectives: This study aimed to determine the international sensitivity index (ISI) for the proposed sixth IS for thromboplastin, human, recombinant (coded 24/114). Methods: An international collaborative study was carried out for the assignment of the ISI to the proposed IS by calibration against the current fifth IS for thromboplastin, human, recombinant. PT measurements were performed by 8 trained operators in 4 laboratories (2 operators in each laboratory) using the harmonized manual tilt tube technique. The World Health Organization calibration procedure was followed. Results: ISI values from 8 operators varied from 1.04 to 1.12, and the mean ISI for the proposed IS coded 24/114 was 1.08 (between-operator coefficient of variation, 2.8%). Conclusion: The material coded 24/114 has been accepted by the World Health Organization as the sixth IS for thromboplastin, human, recombinant. Extensive prior training and external quality assessment surveys for the operators reduced the between-operator coefficient of variation. In this era of metrological traceability, a sustainable reference measurement system has been developed for PT/INR based on the harmonized manual tilt tube technique and a new IS.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
International collaborative study for the calibration of a proposed World Health Organization International Standard for thromboplastin, human, recombinant: a report from the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)—Scientific and Standardization Committee (SSC)/ISTH working group on prothrombin time/international normalized ratio standardization
Date Crossref
01/05/2026
Éditeur
Elsevier BV
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Clinical Laboratory Practices and Quality ControlPlatelet Disorders and TreatmentsVenous Thromboembolism Diagnosis and Management

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.