Development and Dosimetric Characterization of an In Vitro Benchtop Americium 241 Alpha Irradiator Platform
Rattachement africain : us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Purpose To expand preclinical efforts aimed at understanding the unique radiobiology of high linear energy transfer (LET) alpha particles and their implications for cancer therapy, robust, accessible, and high-throughput in vitro alpha irradiation platforms are needed. Such platforms must be extensively characterized and validated to ensure rigor and reproducibility. Here, we describe the construction and dosimetric characterization of a benchtop in vitro alpha irradiation platform using an Americium-241 ( 241 Am) source. Methods and Materials A square foil with embedded 241 Am was used as the alpha source (activity 3.3785 mCi). The source was secured in the roof of a custom housing incorporating a 3D-printed microcapillary array collimator to exclude large-angle alpha particles, thereby narrowing the energy spectrum, and improving homogeneity. A movable shelf accommodates cells grown on glass slides at variable distances from the source. Dosimetric characterization included direct measurement of dose homogeneity, source self-attenuation, energy spectra, absorbed dose, and dose rate. Results The 241 Am source, when used in conjunction with the microcapillary array, provided a spatially uniform and geometrically stable irradiation field. The mylar entrance window resulted in predictable shifting and broadening of the alpha particle energy spectra. Absorbed dose and dose-rate measurements (5.815 - 7.721 mGy/s) demonstrated feasibility of controlled in vitro alpha particle irradiation studies. Conclusions We established a turnkey, low-cost in vitro alpha irradiation platform suitable for high-throughput preclinical studies. The homogeneity, alpha particle energy spectra, absorbed dose and dose rate were characterized and validated through direct measurements. Given the approximately 432-year half-life of 241 Am, this platform will provide stable and reliable dose rates for decades, facilitating long-term radiobiological investigations.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Development and Dosimetric Characterization of an In Vitro Benchtop Americium 241 Alpha Irradiator Platform
- Date Crossref
- 01/06/2026
- Éditeur
- Elsevier BV
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Où se fait cette recherche
-
University of Wisconsin–Madison pays non établi dans la noticeUniversité ou école supérieure
-
The University of Texas MD Anderson Cancer Center Department of Radiation Physics pays non établi dans la noticeÉtablissement de santé
-
Mayo Clinic Departments of Radiation Oncology pays non établi dans la noticeÉtablissement de santé
-
University of Wisconsin-Madison Department of Medical Physics pays non établi dans la noticeUniversité ou école supérieure
-
Departments of Radiation Oncology pays non établi dans la noticeInstitution
-
Radiology pays non établi dans la noticeInstitution
University of Wisconsin–Madison, Department of Radiation Physics — The University of Texas MD Anderson Cancer Center et Departments of Radiation Oncology — Mayo Clinic, avec 3 autres affiliations.
Une affiliation ne permet pas de déduire la nationalité d’un auteur.