Swept-Source OCT Biometry–Derived Corneal Optical Quality: Repeatability Limits for IOL Design Selection in Refractive Cataract Surgery
Rattachement africain : at. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Purpose: To provide the first reliability analysis of swept-source optical coherence tomography (SS-OCT) biometry–derived corneal optical quality metrics that may impact intraocular lens design selection in refractive cataract surgery. Methods: This was a retrospective instrument evaluation study at the Department of Ophthalmology, Hanusch Hospital, Vienna, Austria. Three consecutive preoperative single-operator scans of patients with cataract were analyzed using the updated wavefront analysis software of a SSOCT biometer (ANTERION; Heidelberg Engineering). Corneal higher order aberrations (HOAs) were quantified as paired polar Zernike magnitudes, dioptric vector components, individual signed Zernike coefficients, and root mean square (RMS) values. For each metric, pooled within-subject standard deviation (S w ), coefficient of variation, and intraclass correlation coefficient (ICC) were calculated. Primary analysis was done at a 4-mm optical zone, with additional 3- and 5-mm assessments to evaluate zone-dependent consistency. Results: Fifty-six eyes of 56 patients with cataract were included. All anterior, posterior, and total HOAs revealed repeatability limits ( R = 2.77 × Sw) below 0.10 µm with ICCs of 0.76 to 0.97 and coefficients of variation less than 25% at 4 mm. For the total cornea, RMS HOA and spherical aberration showed repeatability limits of 0.07 µm (ICC: 0.89) and 0.03 µm (ICC: 0.84), respectively. Test–retest variance scaled with zone size (photopic to scotopic) and interindividual variation. Conclusions: Corneal optical quality can be reliably characterized from a SS-OCT biometer at optical zones relevant to refractive cataract surgery. Because measurement uncertainty can impair clinical decision-making, repeatability limits should be considered if wavefront metrics are employed to inform intraocular lens design selection or outcome-based research.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Swept-Source OCT Biometry–Derived Corneal Optical Quality: Repeatability Limits for IOL Design Selection in Refractive Cataract Surgery
- Date Crossref
- 01/06/2026
- Éditeur
- SLACK, Inc.
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.