Bench testing a contact sensing tracheal tube for monitoring the cuff–trachea interface

0Citations signalées — pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

Cuffed tracheal tubes enable a secure airway during mechanical ventilation; however, improper cuff inflation can lead to complications such as inspiratory gas leaks, pulmonary ingress of orogastric secretions, or tracheal injury. This study explores a sensor-based approach to optimise cuff inflation using fibre Bragg grating (FBG) optical sensors embedded within the cuff to detect contact with the tracheal wall. Testing was conducted on multiple trachea models, including cylindrical and bio-inspired models, as well as ex vivo porcine trachea samples, to assess cuff-trachea contact. Seal performance was evaluated at both standard inflation pressures and contact-guided inflation pressures. The contact-sensing tracheal tube successfully detected cuff-trachea contact in all models and ex vivo samples tested. Seal tests showed that larger diameter tracheal models exhibited lower leakage rates at contact-guided inflation pressures compared with standard inflation, indicating improved sealing. In contrast, smaller diameter models showed higher leakage rates at contact-guided pressures due to cuff folding, which formed leakage pathways. These results demonstrate that contact sensing improved cuff performance in larger tracheal models by enhancing sealing effectiveness. The contact-sensing tracheal tube differentiated tracheal model sizes, indicating potential for personalised cuff inflation to accommodate variations in tracheal size and shape. Future work will focus on further sensor miniaturisation and integration within lower-volume cuffs to enable effective contact sensing across a wider range of tracheal geometries and support translation to in vivo use.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Bench testing a contact sensing tracheal tube for monitoring the cuff–trachea interface
Date Crossref
01/06/2026
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Tracheal and airway disordersNosocomial Infections in ICUInhalation and Respiratory Drug Delivery

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.