Aller au contenu principal
Accès ouvert déclaré 2026 article

Comparison of ventilation with second-generation supraglottic airway devices in a prospective randomized cadaver study

0Citations signalées, ce qui n’est pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : de. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Abstract Second-generation supraglottic airway devices (SADs) are widely used alternatives to endotracheal intubation in both anesthesia and emergency airway management. The newly developed LT®evo represents a redesigned version of the established Laryngeal Tube Suction Disposable (LTS-D). This study compared the ventilatory performance of the LT®evo with endotracheal intubation and three other second-generation SADs— LTS-D, Ambu® AuraGain™ and i-gel® Plus—using a controlled human cadaver model. In this prospective, randomized experiment, six thawed adult cadavers were sequentially ventilated via an endotracheal tube (ET) and four SADs under standardized mechanical ventilation (tidal volume 7 mL/kg, PEEP 5 cmH₂O, respiratory rate 10/min). Airway pressure and flow were continuously recorded (sampling rate 200 Hz) to determine inspiratory tidal volume, pressure, flow, and calculated resistance. Endotracheal intubation using fiberoptics was attempted via LT®evo , AuraGain™ and i-gel® Plus. Data were analyzed using Kruskal–Wallis tests with Holm-corrected post-hoc pairwise comparisons. From six human cadavers (60–85kg estimated body weight), 30 complete ventilation datasets were obtained. Initial fiberoptic control of SAD placement showed optimal positioning in 13 of 24 insertions. During ventilation cycles, inspiratory tidal volumes of the LT®evo (0.49 [0.46–0.52]) were similar to those of ET, AuraGain™ and LTS-D and superior to those of the i-gel® Plus (0.31 L [0.27–0.33], p < 0.001). Peak inspiratory pressure and inspiratory resistance were higher in ET when compared to LT®evo and all other SADs. Endotracheal intubation using fiberoptics was successful in all attempts via LT®evo , AuraGain™ and i-gel® Plus. In this controlled cadaver study, the newly designed LT®evo demonstrated ventilatory performance comparable to that of the LTS-D and the Ambu® AuraGain TM and was superior to the i-gel® Plus in delivering adequate tidal volumes. Additionally, LT®evo allowed for fiberoptic ET placement. Importantly, these results are specific to the cadaveric model and reflect technical variability in ventilation mechanics rather than clinical effectiveness. Clinical trial number German Clinical Trials Registry, DRKS00038309, on October 30 th , 2025.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Comparison of ventilation with second-generation supraglottic airway devices in a prospective randomized cadaver study
Date Crossref
21/05/2026
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Où se fait cette recherche

  • Heidelberg University pays non établi dans la notice
    Université ou école supérieure
  • University Hospital Heidelberg pays non établi dans la notice
    Établissement de santé
  • Medical Faculty Heidelberg pays non établi dans la notice
    Université ou école supérieure

Heidelberg University, University Hospital Heidelberg et Medical Faculty Heidelberg.

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Airway Management and Intubation TechniquesNosocomial Infections in ICURespiratory Support and Mechanisms

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.