Data-Driven Filter for Detector Oscillation Artifacts in Time-of-Flight Mass Spectrometry

0Citations signalées — pas une note de qualité
4Institutions déclarées
3Pays d’affiliation déclarés

Résumé fourni par la source

High Resolution Image Download MS PowerPoint Slide Time-of-flight mass spectrometry (TOF-MS) is widely used for complex mixture analysis due to its ability to detect thousands of chemical compounds in a single sample. However, high-intensity ions can saturate the detector and generate artifacts, commonly referred to as “detector ringing”. These artifacts resemble true MS features and often persist after componentization, complicating data interpretation and feature prioritization. Here, we introduce a data-driven filtering approach that exploits the time-domain origin of detector ringing to remove such artifacts. Because ion flight times scale with m / z, oscillations in the detector response manifest as artifactual features with constant delays in m / z, providing a universal detection criterion independent of ion source conditions and chromatographic or ion mobility separation. The workflow was evaluated on GC-, LC-, and ion mobility-hyphenated TOF-MS datasets from multiple instrumental platforms, identifying and removing up to 5.3% of features. Comparison across instruments revealed that Δ m / z spacings are not only constant but also instrument-specific and polarity-independent, suggesting potential applications in data forensics and traceability. To assess broader relevance, the filter was applied to the MassBank and NIST DART-MS Forensics fragmentation mass spectral libraries. Although the overall incidence was low relative to library size (1.7%), over 1000 mass spectra were provisionally identified as containing ringing artifacts. Clustering these spectra by their Δ m / z effectively partitioned the data by TOF-MS platform, revealing distinct groupings across libraries and contributors. By specifically targeting detector-derived artifacts, this workflow reduces false positives and improves feature prioritization and compound identification, enhancing the robustness of TOF-MS-based chemical analyses.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Data-Driven Filter for Detector Oscillation Artifacts in Time-of-Flight Mass Spectrometry
Date Crossref
12/05/2026
Éditeur
American Chemical Society (ACS)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Mass Spectrometry Techniques and ApplicationsAnalytical chemistry methods developmentSpectroscopy and Laser Applications

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.