Clinical evaluation of two new unshielded and shielded silicon diode detectors in external photon beam radiation therapy: SunSILICON and SunSILICON P

3Citations signalées, ce qui n’est pas une note de qualité
7Institutions déclarées
4Pays d’affiliation déclarés

Rattachement africain : de, us, pl, fr. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: Advanced external photon beam radiotherapy requires dosimetry detectors with high spatial resolution and minimal perturbation effects. Silicon diode detectors are widely used due to their high sensitivity and small sensitive volumes, but fluence perturbation effects and energy dependence can affect performance and depend on detector construction. Shielded and unshielded configurations are used to address these challenges, but their clinical performance and limitations require systematic evaluation. PURPOSE: This study aims to evaluate the clinical performance of two newly developed silicon diode detectors-SunSILICON (unshielded) and SunSILICON P (shielded)-for relative dosimetry in external photon beam radiotherapy. The detectors' response behaviors are compared to established detectors across the full clinical range of field sizes and beam energies, focusing on percentage depth dose (PDD) and lateral beam profile measurements. METHODS: Co unit, using various types of motorized water phantoms. Comparative measurements were performed with ionization chambers, as well as other silicon and diamond detectors. Over 600 scans were analyzed, and gamma analysis was applied to assess agreement using 0.5%/0.5 mm for PDDs and 1%/1 mm for lateral beam profiles. Angular dependence, detector sensitivity and the effective point of measurement (EPOM) were determined in solid phantoms. RESULTS: SunSILICON and SunSILICON P demonstrated excellent agreement with established detectors within their specified field size ranges. Gamma passing rates exceeded 99% for most comparisons, with minor deviations in large fields or at field size limits. The shielded SunSILICON P showed reduced energy dependence in large fields compared to the unshielded version, while SunSILICON showed superior performance in small fields. Both detectors exhibited minimal angular dependence (<1% for clinically relevant angles) and negligible intra-type variation. The measured EPOMs matched calculated values within uncertainty. The combination of both detectors enables comprehensive coverage of clinical dosimetry needs. CONCLUSIONS: SunSILICON and SunSILICON P provide reliable, high-resolution dosimetry for external photon beam radiotherapy across a broad range of clinical scenarios. Their performance is comparable to or exceeds that of established silicon and diamond detectors, with the shielded version particularly suited for larger fields due to its reduced energy dependence. Together, these detectors offer a robust solution for clinical relative dosimetry.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Clinical evaluation of two new unshielded and shielded silicon diode detectors in external photon beam radiation therapy: SunSILICON and SunSILICON P
Date Crossref
01/04/2026
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Advanced Radiotherapy TechniquesRadiation Therapy and DosimetryRadiation Detection and Scintillator Technologies

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.