Gradient system characterization of a 1.5 T MR-Linac with application to 4D UTE imaging for adaptive MR-guided radiotherapy of lung cancer

0Citations signalées, ce qui n’est pas une note de qualité
0Institutions déclarées
0Pays d’affiliation déclarés

Le résumé fourni par la source

Purpose: To measure the gradient system transfer function (GSTF) of an MR-Linac (Elekta Unity, Stockholm, Sweden) using an accessible phantom-based method and to apply trajectory corrections for UTE image reconstruction in the context of MR-guided radiotherapy of lung cancer. Methods: The first-order GSTF of a 1.5 T, split gradient Elekta Unity MR-Linac was measured using a thin-slice technique to characterize gradient system imperfections for each physical gradient axis (X, Y, Z). Repeatability measurements of the GSTF were performed 48 h apart. The GSTF was applied to trajectory correction in multi-echo UTE image reconstruction (TEs = 0.176, 1.85, 3.52 ms) to allow for UTE-Dixon inputs in the generation of synthetic CT. Images were acquired in an anthropomorphic phantom and in two free-breathing lung cancer patients. For patient scans, respiratory-correlated 4D-MR images were reconstructed using self-navigation and an iterative compressed-sensing algorithm. Results: The GSTF magnitude was similar across the X/Y/Z axes up to ˜6 kHz. The GSTF phase was similar between the X/Y/Z components up to ˜3 kHz. Repeatability measurements demonstrated minimal variations corresponding to a system delay difference of 0.06 μs. Corrected UTE trajectory spokes are shifted approximately 1 m−1 compared to the nominal k-space location. Corrected phantom and patient UTE images exhibited improved signal uniformity and contrast and reduced halo and signal loss artifacts. Trajectory correction for the later TE images did not improve overall image quality. Conclusion: The proposed GSTF measurement method using standard MR-Linac hardware enables successful trajectory correction in UTE imaging reconstruction, with applications to lung synthetic CT generation for MR-guided radiotherapy.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

Aucun DOI disponible pour le contrôle Crossref.

Les sujets associés

Advanced Radiotherapy TechniquesAdvanced MRI Techniques and ApplicationsRadiation Therapy and Dosimetry

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.