X-ray Response of the Fully-Depleted, p-Channel SiSeRO-CCD
Rattachement africain : us, ar. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
We present an X-ray characterization of a fully depleted, 725 $μ$m thick p-channel SiSeRO-CCD. Measurements with a $^{55}$Fe source yield an energy resolution of $54 \pm 0.9$ eV ($14.6 \pm 0.25 e^{-}$) at 5.9 keV for single-pixel events, indicating that the SiSeRO amplifier preserves the intrinsic charge resolution of the CCD under multi-sample non-destructive readout. Characterization with a $^{241}$Am source extends the response to higher-energy photons, with reconstructed spectral features observed between 9-26 keV and the 59.5 keV $γ$ emission. These measurements, together with a muon-derived diffusion calibration, show that charge transport and diffusion are consistent with interactions spanning the full sensor depth. These results demonstrate that the SiSeRO-CCD simultaneously achieves sub-electron noise performance and efficient charge collection in a thick, fully depleted silicon detector. This combination enables X-ray spectroscopy across a broad energy range while maintaining sensitivity to faint signals.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.