Aller au contenu principal
2026 article

Monte Carlo–based optimization of CTV‐to‐PTV margins for image‐guided VMAT prostate radiotherapy

0Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: The van Herk margin formula, derived for 3-D conformal radiotherapy with a uniform 3.2 mm dose penumbra and no intrinsic dose buffer, remains widely used to design CTV-to-PTV expansions in contemporary image-guided prostate radiotherapy. However, coplanar VMAT techniques often feature broader penumbrae and explicit PTV-isodose clearances, potentially violating the assumptions underlying the original formulation and leading to overly conservative margins. PURPOSE: We investigated whether the widely adopted van Herk margin formula overestimates clinical target volume (CTV)-to-planning target volume (PTV) expansions for contemporary coplanar volumetric modulated arc therapy (VMAT) prostate treatments. METHODS: Fifty consecutive intact-prostate VMAT radiotherapy plans (two coplanar arcs; clinical margins 3 mm, except for 2 mm posterior) were exported. Direction-specific 90% isodose-to-PTV gaps and penumbra widths were measured. Candidate anisotropic margins were tested by eroding the PTV to create CTV_eval. Monte-Carlo simulations combined systematic (Σ) shifts with Gaussian random (σ') blurring kernels of 0-2 mm were performed. Acceptability criteria of (i) CTV Dmin0.03 cc ≥ 90% Rx in ≥ 90% of simulated scenarios or (ii) population tumor-control probability (TCP) loss < 1 % were used. RESULTS: VMAT plans exhibited intrinsic 90% isodose clearances of 3-5 mm laterally/anteriorly and 1-2 mm superior-inferiorly, while axial-plane penumbras were up to fivefold broader than van Herk's assumption. With a [0, 0, 2] mm (LR/AP/SI) margin, ≥ 90% of patients maintained Dmin0.03 cc ≥ 90% Rx provided Σ lay within an ellipsoid of [2.0, 1.5, 1.8] mm and σ' ≤ [1.5, 2.0, 1.5] mm. Under TCP criteria the safe Σ ellipsoid for high-risk disease was [2.5, 1.9, 2.2] mm, while low-intermediate risk was even less sensitive. CONCLUSIONS: For image-guided coplanar VMAT prostate radiotherapy, an anisotropic [0, 0, 2] mm CTV-to-PTV margin is sufficient for target coverage. A modified margin expression that subtracts the measured isodose-to-PTV gap and uses reduced random-error coefficients better reflects modern practice.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Monte Carlo–based optimization of CTV‐to‐PTV margins for image‐guided VMAT prostate radiotherapy
Date Crossref
30/03/2026
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Advanced Radiotherapy TechniquesProstate Cancer Diagnosis and TreatmentRadiation Therapy and Dosimetry

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.