Manual Insertion of Cochlear Implant Electrodes Versus Robot-Assisted Insertion and Analysis by Micro-CT: A Temporal Bone Study
Résumé fourni par la source
Background/Objectives: Atraumatic electrode array insertion should be targeted in cochlear implantation. Robotic insertion is used in many centers worldwide. Our objective was to evaluate manual electrode placement and robot-assisted placement using RobOtol® on human temporal bones (TBs), in terms of endocochlear trauma and completion of insertion. Methods: Sixteen TBs originating from eight bodies were implanted with Medel-FLEX24 electrodes through the round window. The right TB was implanted manually, while the left TB of the same body was implanted using RobOtol® for electrode insertion. Results were analyzed through micro-computed tomography imaging. No statistical analysis was used, given the small sample size; a descriptive interpretation of micro-CT scans was rather preferred. Results: In the “manual group”, there were two cases (25%) of insertion trauma: elevation of basilar membrane at basal turn (Eshraghi-stage-1). In the “robotic group”, there were two cases (25%) of insertion trauma: one case of elevation of basilar membrane at the middle turn (Eshraghi-stage-1) and one case of dislocation of all electrodes in scala vestibuli (Eshraghi-stage-3). There were six cases (75%) of incomplete insertion in the “manual group” and four cases (50%) of incomplete insertion in the “robotic group”. Conclusions: Both techniques of electrode placement yielded fairly similar results, in terms of endocochlear trauma and completion of insertion. New larger-scale cadaveric and clinical studies are needed to determine the possible benefit of robot-assisted electrode insertion in cochlear implantation.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Manual Insertion of Cochlear Implant Electrodes Versus Robot-Assisted Insertion and Analysis by Micro-CT: A Temporal Bone Study
- Date Crossref
- 26/03/2026
- Éditeur
- MDPI AG
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.
Institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.