Point-of-Care Bilirubin Testing in Neonates: Comparative Performance of Blood Gas Analysis and Transcutaneous Bilirubinometry
Résumé fourni par la source
Background: Neonatal jaundice is a common condition with potentially severe complications such as bilirubin-induced neurological dysfunction and kernicterus. While serum bilirubin (SBR) remains the standard laboratory measurement, point-of-care methods, such as transcutaneous bilirubinometry (TcB) and blood gas analysers (BGAs), offer rapid, less invasive alternatives. Direct comparisons of their diagnostic accuracy remain limited. Objective: The aim of this study was to assess and compare diagnostic accuracy and clinical utility of TcB and BGA against SBR in neonatal hyperbilirubinaemia screening. Methods: This retrospective study included neonates (n = 221) with concurrent SBR, BGA, and TcB measurements (n = 333). Assessment was via Passing–Bablok regression, Bland–Altman analysis, and Spearman correlation. Diagnostic performance was evaluated against jaundice thresholds in phototherapy charts (≥95th percentile threshold). Subgroup analyses considered phototherapy status, haemoglobin concentration, and Fitzpatrick skin type. Results: BGA showed stronger agreement with SBR (R2 = 0.88) than TcB (R2 = 0.43). BGA remained accurate regardless of phototherapy or haemoglobin levels. TcB accuracy declined post-phototherapy with reduced predictive value in darker-skinned neonates (Fitzpatrick III–VI) and increased false discovery rates. Both methods demonstrated low sensitivity (45.8%) but high specificity (>95%) and negative predictive value (~91%) for clinically significant hyperbilirubinaemia. BGA had a higher diagnostic odds ratio (47.5) than TcB (19.3). When individual patient sequential SBR and BGA measurements were compared for jaundice tracking (n = 175), there was high correlation, (r = 0.971) with no statistical differences, and 50% of measurements achieving agreement within 10 μmol/L. Conclusions: BGA is a more reliable alternative to SBR than TcB, particularly in time-critical or resource-limited settings. While TcB remains a non-invasive screening tool, limited accuracy post-phototherapy and with darker skinned neonates indicate confirmatory SBR testing. These findings support the selective and context-aware use of BGA and TcB to optimise neonatal hyperbilirubinaemia management and reduce interventions.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Point-of-Care Bilirubin Testing in Neonates: Comparative Performance of Blood Gas Analysis and Transcutaneous Bilirubinometry
- Date Crossref
- 01/02/2026
- Éditeur
- MDPI AG
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.
Institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.