Abstract TP151: 30-Minute ABPM Underdetects White Coat Hypertension Despite Good Diagnostic Accuracy

0Citations signalées — pas une note de qualité
12Institutions déclarées
4Pays d’affiliation déclarés

Résumé fourni par la source

Introduction: Hypertension is the leading modifiable stroke risk factor, yet 24-hour ambulatory blood pressure monitoring (ABPM) access is limited by patient burden and capacity constraints. Shorter 30-minute protocols may expand population-based stroke screening but could misclassify white coat hypertension (WCH), causing unnecessary treatment and inefficient resource allocation. Aims: To compare 30-minute versus 24-hour ABPM for diagnosing hypertension and detecting WCH, and evaluate a tiered approach for stroke prevention in resource-limited settings. Methods: Cross-sectional study of 90 Vietnamese adults from routine care (median age 34 years [IQR 25-55]; 60% male) completing both protocols. Office hypertension was ≥140/90 mmHg. Normal ambulatory BP was <130/80 mmHg (24-hour) and <135/85 mmHg (30-minute). Agreement assessed using Bland-Altman analysis; diagnostic performance using ROC curves; classification consistency using Cohen's kappa. Results: Mean BP by 30-minute monitoring was 136.0±13.1/84.1±11.7 mmHg versus 130.4±12.6/77.5±9.3 mmHg by 24-hour ABPM. Strong correlations existed for SBP (r=0.687, R2=0.479) and DBP (r=0.787, R2=0.620), both p<0.001. Bland-Altman analysis revealed systematic bias with 30-minute ABPM overestimating BP by +5.5 mmHg for SBP (95% LoA: -14.3 to 25.4 mmHg) and +6.7 mmHg for DBP (95% LoA: -7.4 to 20.8 mmHg), indicating substantial individual variability. ROC analysis demonstrated good diagnostic performance: SBP AUC=0.830 with 94.5% sensitivity and 62.9% specificity; DBP AUC=0.786 with 80.0% sensitivity and 77.1% specificity. Overall classification accuracy was 78.9% with moderate agreement (κ=0.544). For WCH detection, 24-hour ABPM identified 27/90 patients (30.0%) versus 20/90 (22.2%) by 30-minute ABPM, representing 7.8% absolute underdetection rate. Conclusions: Thirty-minute ABPM provides accessible screening with acceptable accuracy but systematically overestimates BP and underdetects WCH. For population-level stroke prevention, we recommend a two-step pathway: first-line 30-minute monitoring with interpretive adjustments (subtract 5–7 mmHg), followed by confirmatory 24-hour monitoring for positive screens. This tiered approach prioritizes accurate risk stratification, prevents WCH overtreatment, and enables cost-effective stroke prevention in resource-constrained health systems.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Abstract TP151: 30-Minute ABPM Underdetects White Coat Hypertension Despite Good Diagnostic Accuracy
Date Crossref
01/02/2026
Éditeur
Ovid Technologies (Wolters Kluwer Health)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Blood Pressure and Hypertension StudiesCardiovascular Health and Risk FactorsAcute Ischemic Stroke Management

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.