The effects of measurement errors on minimum ablative margins after thermal ablation of liver tumors: a simulation study
Rattachement africain : us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Purpose: To develop a mathematical framework to estimate the in silico A0 threshold based on the technical specifications of a specific ablation confirmation software package for thermal ablation of liver tumors that can then be used to identify the impact of different sources of error. Methods: To estimate in silico A0 thresholds, we developed a simulation framework incorporating technical parameters and biological effects. Technical parameters were segmentation error, registration error, and slice thickness, and biological effects were tissue shrinkage and microscopic satellite lesions; these parameters and effects were all modeled using statistical distributions. For each permutation of parameters, a logistic regression was fitted to determine the observed MAM required to achieve ≥ 99% probability of true complete tumor coverage (i.e., the A0 threshold). The mathematical framework was integrated into a web application to estimate the A0 threshold and the reliability of the commonly used 5-mm A0 threshold based on several software performance characteristics. Results: A total of 15,000,000 simulations (10,000 simulations × 1500 parameter permutations) were run and summarized. Tumor and ablation zone segmentation most greatly influenced the A0 threshold, with thresholds of 3.4 and 8.4 mm for 1- and 5-mm errors, whereas slice thickness had a relatively small effect, with A0 thresholds of 2.9 and 3.4 mm for thicknesses of 1 and 5 mm, respectively. Conclusion: This framework provides a method to determine software-specific in silico A0 thresholds and evaluate the reliability of existing 5-mm criteria based on software performance metrics. The results further show that ablation confirmation software should have registration and segmentation errors of ≤ 3 mm to reliably use a 5-mm A0 threshold.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- The effects of measurement errors on minimum ablative margins after thermal ablation of liver tumors: a simulation study
- Date Crossref
- 14/01/2026
- Éditeur
- Springer Science and Business Media LLC
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.