Monopolar biphasic focal pulsed field ablation directly at the atrioventricular junction and from within the noncoronary cusp: The PFA-CONDUCT study
Résumé fourni par la source
BACKGROUND: The efficacy and safety of monopolar biphasic focal pulsed field ablation (F-PFA) in close proximity to the atrioventricular (AV) node remain unknown. OBJECTIVE: This study aimed to describe the effect of direct or indirect F-PFA application at the AV junction. METHODS: In pigs, F-PFA (Centauri, CardioFocus) was applied directly at the AV junction (n = 3) and indirectly from within the aortic noncoronary cusp (NCC) (n = 5), followed by macroscopic gross and histologic analysis. In 5 patients planned for a pace-and-ablate strategy, F-PFA was applied at the AV junction. In 1 patient with recurrent para-Hisian atrial tachycardia and previously failed ablation, F-PFA was applied from within the NCC. RESULTS: In pigs, direct F-PFA applications at the AV junction were associated with junctional beats and resulted in complete AV block after 2-3 F-PFA applications. Tissue histology revealed a lesion depth of approximately 8 mm with minor hemorrhage and leukocyte infiltration. F-PFA from within the NCC resulted in a transient increase in the AV node's Wenckebach cycle length and 1 pig had transient AV block, which recovered within 10 minutes. In 5 patients undergoing pace-and-ablate strategy, F-PFA applications at the AV junction resulted in sustained AV block. In 1 patient, application of F-PFA from within the NCC resulted in termination and noninducibility of a recurrent incessant para-Hisian atrial tachycardia without AV block. There were no immediate or long-term complications. CONCLUSION: F-PFA directly at the AV junction results in nonreversible AV block. F-PFA from within the NCC may represent a promising strategy for treating para-Hisian arrhythmias while preserving AV conduction.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Monopolar biphasic focal pulsed field ablation directly at the atrioventricular junction and from within the noncoronary cusp: The PFA-CONDUCT study
- Date Crossref
- 01/04/2026
- Éditeur
- Elsevier BV
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.
Institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.