Ex Vivo Quantitative Evaluation of Beam Hardening Artifacts at Various Implant Locations in Cone-Beam Computed Tomography Using Metal Artifact Reduction and Noise Reduction Techniques

0Citations signalées, ce qui n’est pas une note de qualité
5Institutions déclarées
4Pays d’affiliation déclarés

Rattachement africain : tr, dk, ru, cn. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Purposes: Beam hardening artifacts caused by dental implants remain one of the most significant limitations of cone-beam computed tomography (CBCT), often compromising the evaluation of peri-implant bone and potentially masking critical diagnostic findings. Although metal artifact reduction (MAR) and noise-optimization filters such as the Adaptive Image Noise Optimizer (AINO) are widely available in commercial CBCT systems, their effectiveness varies depending on implant configuration and scanning parameters. A clearer understanding of how implant positioning influences artifact severity—together with how MAR and AINO perform under different conditions—is essential for improving diagnostic reliability. Materials and Methods: A fresh frozen cadaver head, with dental implants inserted using two configurations (C1 and C2), was scanned using different scan parameters, with and without metal artifact reduction and image optimization filters. The percentages of gray value alteration due to artifacts were evaluated, using registered pre-implant scans as a control. Regions of interest were defined by an experienced researcher. For the two implant conditions, ROIs were placed as follows: C1—lingual, buccal and mesial to the mesial implant; lingual, buccal and distal to the distal implant; and an additional ROI between the implants (n = 7); C2—lingual, buccal, mesial and distal to each implant (n = 8). For each ROI, the mean gray value was measured in five consecutive axial slices, and rescaled according to calibration points in air and soft tissue. Results: Significant differences were found in gray values across configurations and scan modes. In the C2 configuration, combined MAR and AINO restored gray values in certain ROIs from 1.227 (OFF) to 1.223 (MAR+AINO), closely matching the control (1.227). In contrast, C1 showed limited improvement; for example, buccal ROI gray values decreased from 3.978 (OFF) to 3.323 (AINO) compared to the control (3.273), with no significant benefit from additional MAR. Conclusion: Artifacts from implants can be significantly affected by their (relative) position and the use of MAR and AINO.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Ex Vivo Quantitative Evaluation of Beam Hardening Artifacts at Various Implant Locations in Cone-Beam Computed Tomography Using Metal Artifact Reduction and Noise Reduction Techniques
Date Crossref
15/12/2025
Éditeur
MDPI AG
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Dental Radiography and ImagingAdvanced X-ray and CT ImagingMedical Imaging Techniques and Applications

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.