Diagonal DWI : A Time‐Efficient Alternative to 3‐Scan Trace DWI for Breast Lesion Evaluation at 3.0‐T MRI —A Phantom Study and Clinical Assessment
Rattachement africain : jp, de. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
BACKGROUND: Diffusion-weighted imaging (DWI) in breast MRI requires balancing image quality with acquisition time. Reducing scan time without sacrificing diagnostic performance could improve patient comfort and workflow. PURPOSE: To compare 3-scan trace DWI (tDWI) and diagonal DWI (dDWI) for breast lesion evaluation using both phantom and clinical assessments, focusing on image quality metrics and diagnostic performance. STUDY TYPE: Retrospective comparative study. POPULATION: A commercially available breast diffusion phantom (Caliber MRI; Boulder, CO, USA) and 92 consecutive participants were initially enrolled. After excluding 23 due to having no confirmed lesion, and 15 with non-mass lesions, 54 patients (median age 55.5 years, range 21-82) were analyzed. FIELD STRENGTH/SEQUENCE: 3-T, tDWI (120 s) and dDWI (74 s) with identical parameters except for diffusion gradient directions, with dDWI reducing scan time by 38%. ASSESSMENT: Phantom studies measured apparent diffusion coefficient (ADC) and estimated signal-to-noise ratio (eSNR). Clinical studies evaluated ADC, contrast-to-noise ratio (CNR), and contrast ratio (CR) from ROIs in lesions, normal breast tissue, and fat. Three radiologists scored lesion conspicuity and image quality. STATISTICAL TESTS: Due to non-normal data distribution, the Wilcoxon signed-rank test compared metrics between sequences. The ADC coefficient of variation (CV) was calculated. A p-value < 0.05 was considered significant. RESULTS: ). tDWI demonstrated significantly higher CNR for breast tissue and fat. tDWI also showed lower CR values for tumor/breast tissue and lower values for tumor/fat. ADC measurements were similar between techniques (CV = 4.20%). DATA CONCLUSION: dDWI provides a 38% shorter acquisition time than tDWI while maintaining comparable quantitative performance. dDWI demonstrates improved image quality, particularly in challenging anatomical regions, though tDWI yields higher contrast-to-noise ratio values. EVIDENCE LEVEL: 3. TECHNICAL EFFICACY STAGE: 2.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé, mais le titre doit être comparé manuellement.
- Titre Crossref
- Diagonal <scp>DWI</scp> : A Time‐Efficient Alternative to 3‐Scan Trace <scp>DWI</scp> for Breast Lesion Evaluation at 3.0‐T <scp>MRI</scp> —A Phantom Study and Clinical Assessment
- Date Crossref
- 12/12/2025
- Éditeur
- Wiley
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.