Fast and accurate measurement of small field dosimetry using a novel scintillation detector

0Citations signalées, ce qui n’est pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Abstract Background The most used instruments for small‐field dosimetry have notable limitations, including the need for correction of output factors, limited scanning speeds, and challenges in alignment for percentage depth dose (PDD) measurements, particularly for extremely small fields. However, plastic scintillation detectors (PSDs) are an attractive alternative for small‐field dosimetry due to their correction‐free nature, linear dose response, and fast response time. Purpose This study evaluates the robustness and accuracy of the dosimetric measurements using a new water‐equivalent PSD in small‐field dosimetry. The study also aims to report accurate measurements of output factors, profiles, and an indirect method for measuring PDD in small fields, with a scanning time that is 5 to 10 times faster than traditional methods. Method PDDs, profiles, and output factors were measured on a Varian TrueBeam 6XFFF photon beam for the field sizes of 0.5 × 0.5 cm 2 , 1 × 1 cm 2 , 2 × 2 cm 2 , 3 × 3 cm 2 , and 4 × 4 cm 2 using a new PSD from blue physics (BP‐PSD). These measurements were compared with those obtained using a well‐established PSD (Standard Imaging W2), micro‐diamond (TN60019, PTW‐Freiburg, Germany), and micro‐silicon detectors (TN60023, PTW‐Freiburg, Germany). Owing to its fast response, the BP‐PSD enabled the collection of beam profiles at 31 depths, which were used to derive the PDD while avoiding detector misalignment along the beam path. Data was collected in a water tank controlled by the PTW BeamScan software. The pulse‐by‐pulse raw data from BP‐PSD were converted to respective dosimetry data using in‐house software. Result The BP‐PSD demonstrated excellent agreement with other detectors for small‐field output factors (FOFs), with a maximum variation of 1.6%. The BP‐PSD also showed strong agreement in PDD measurements with an ion chamber (TN31013) for both 3 × 3 cm 2 and 10 × 10 cm 2 field sizes, achieving a 98% gamma passing rate (gamma criteria: 1 mm, 3%). For the profile measurements, the BP‐PSD showed consistency with both the micro‐diamond and micro‐silicon diode detectors, with less than 1% variation in measured penumbra length. At a 3 × 3 cm 2 field size, the measured penumbra length (4 mm) agreed with previously published data (3.86–4.2 mm). Additionally, for an extremely small field size (0.5 × 0.5 cm 2 ), the indirect PDD measurements derived from profiles showed significant improvement compared to the direct measurement without beam inclination correction. Conclusion The BP‐PSD has proven to be a robust and reliable detector for small‐field dosimetry. It shows excellent agreement with other detectors in measuring small FOFs, profiles, and PDDs, while achieving significantly faster scanning speeds in a water tank.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Fast and accurate measurement of small field dosimetry using a novel scintillation detector
Date Crossref
30/11/2025
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Radiation Effects and DosimetryRadiation Detection and Scintillator TechnologiesRadiation Therapy and Dosimetry

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.