Computed tomography-based radiostereometric analysis achieves sufficient precision for femoral implants in total knee arthroplasty: a randomized controlled trial

0Citations signalées, ce qui n’est pas une note de qualité
6Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : no. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: Conventional radiostereometric analysis (RSA) has been used for several decades to assess implant migration in total knee arthroplasty (TKA). More recently, computed tomography-based radiostereometric analysis (CT-RSA) has been introduced as an alternative. Experimental studies have demonstrated acceptable precision for CT-RSA; however, this has not yet been validated in a clinical setting. With the emergence of new alignment philosophies in TKA, migration analysis of femoral components has become increasingly important. A key criticism of CT-RSA has been the higher radiation doses compared with conventional RSA, particularly in longitudinal studies involving repeated CT scans. In this study, we evaluated the precision of CT-RSA for assessing femoral component migration in two TKA designs at two different radiation dose levels. We hypothesized that there would be no clinically relevant difference in precision between the two radiation dose levels. METHODS: We performed a randomized controlled trial of 50 patients using Tritanium and GMK Sphere 3D Metal implants and analysed the precision of the method through double examinations expressed as means with 95% confidence intervals at standard and low-dose levels. The main outcome variable was the difference in mean maximum total point motion. We set the clinical meaningful difference to be more than 0.1 mm based on existing literature. Secondary outcome variables were the means and standard deviations of the translation and rotation of the implants. FINDINGS: We found a precision of (mean, standard deviation) 0.20 (0.08) and 0.17 mm (0.06) for standard and low dose respectively, giving a difference (95% confidence interval) of - 0.03 mm (- 0.06-0.00). P-value was 0.031. Variability ratios were 0.009 for the standard deviation test. INTERPRETATION: In this first clinical study to assess the precision of computed tomography-based radiostereometric analysis for femoral implants in total knee arthroplasty, we found no clinically relevant difference in precision. Our findings confirm that CT-RSA provides sufficient precision in a clinical setting to enable migration analysis of femoral implants in total knee arthroplasty, even when using very low radiation doses.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Computed tomography-based radiostereometric analysis achieves sufficient precision for femoral implants in total knee arthroplasty: a randomized controlled trial
Date Crossref
28/11/2025
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Total Knee Arthroplasty OutcomesOrthopaedic implants and arthroplastyAdvanced Radiotherapy Techniques

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.