Direct determination of nickel content in high-salt liquids by EDXRF
Rattachement africain : au, cn. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
For rapid Energy Dispersive X-ray Fluorescence (EDXRF) analysis of High-grade Nickel leachate in hydrometallurgy of Ni ore, a device suitable for detecting highly corrosive liquids and a corresponding EDXRF method for high-concentration Ni solutions were developed. Results show: in the device, NiKα intensity in Ni(NO3)2 mixtures exhibits a parabolic trend with increasing Ni content; at a specific “x-ray source-detector” distance, Ni content corresponding to maximum NiKα intensity is 2.0%~3.0%; for industrial samples with 6.58%-7.93% Ni, NiKα intensity correlates negatively with Ni content, converting to a positive linear correlation after scatter internal standard correction; via the empirical coefficient method, EDXRF analysis of Ni shows absolute error ≤0.09% and repeated analysis RSD ⪅0.6%.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Direct determination of nickel content in high-salt liquids by EDXRF
- Date Crossref
- 21/11/2025
- Éditeur
- SPIE
- Type
- proceedings-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.