A linear model‐based, B 1 inhomogeneity–corrected, glutamate‐weighted chemical exchange saturation transfer MRI at 5 T
Rattachement africain : cn. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Abstract Purpose Glutamate‐weighted chemical exchange saturation transfer (gluCEST) MRI shows promise in the diagnosis and evaluation of neurological disorders. GluCEST must be conducted at high magnetic field strengths, in which pronounced B 1 inhomogeneity compromises gluCEST measurements. Conventional B 1 inhomogeneity correction methods rely on interpolation algorithms, B 1 choices, acquisition numbers, or calibration curves, making correction unrobust. This study proposes a linear model‐based method for B 1 inhomogeneity correction in gluCEST MRI. Methods Four healthy volunteers and 4 brain‐tumor patients underwent B 1 field mapping and chemical exchange saturation transfer (CEST) imaging at three nominal B 1 levels of 3.5, 4.0, and 4.5 μT at 5 T. The linear relationship between 1/Z(±3 ppm) and B 1 was investigated using numerical simulations and human‐brain data. The 1/Z(±3 ppm) values at nominal B 1 were calculated based on the linear function constructed from either two or three CEST acquisitions. GluCEST signals, quantified with 1‐Z(+3 ppm)/Z(−3 ppm), were compared before and after B 1 inhomogeneity correction, including those corrected with the conventional interpolation method. Results A linear dependency of 1/Z(±3 ppm) on B 1 within a B 1 range of 2–5 μT was identified. Evident B 1 inhomogeneity artifacts were observed in the uncorrected gluCEST images. GluCEST signals corrected using the conventional interpolation method varied with CEST acquisition numbers, particularly in regions with severe B 1 inhomogeneity. In contrast, the linear model yielded consistent gluCEST contrasts with less dependence on CEST acquisition numbers, superior to the interpolation method. Conclusion The proposed linear model allows robust B 1 inhomogeneity correction from at least two CEST acquisitions, providing an effective way for improved gluCEST MRI.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé, mais le titre doit être comparé manuellement.
- Titre Crossref
- A linear model‐based, <scp> B <sub>1</sub> </scp> inhomogeneity–corrected, glutamate‐weighted <scp>chemical exchange saturation transfer MRI</scp> at <scp>5 T</scp>
- Date Crossref
- 12/10/2025
- Éditeur
- Wiley
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.