Shunt Dysfunction Assessment in Shunted Patients via Multiparametric Non-Invasive ICP Waveform Monitoring

0Citations signalées, ce qui n’est pas une note de qualité
9Institutions déclarées
2Pays d’affiliation déclarés

Rattachement africain : br, us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Background/Objectives: Ventriculoperitoneal shunts are the standard treatment for adults diagnosed with several CSF disorders, but often face dysfunction, leading interest in non-invasive methods for diagnosing shunt issues. This study evaluates the potential of non-invasive intracranial pressure waveform monitoring (nICPw) with the brain4care (B4C) system to distinguish overdrainage, underdrainage, and normal shunt function in patients with CSF disorders. Methods: In this single-center, observational study at Hospital das Clínicas, Brazil, adult patients with CSF shunts were enrolled. Patients were categorized as Overdrainage or Underdrainage, based on clinical parameters, with an Asymptomatic group. The B4C system provided nICPw monitoring, and six parameters (including various P2/P1 ratios) were analyzed via MANOVA and ANOVA. Results: Among 30 patients (6 overdrainage, 6 underdrainage, 18 asymptomatic), five asymptomatic patients were excluded from the main analysis due to incomplete data collection, leaving 25 patients. Overdrainage patients had significantly higher ΔP2/P1 values (0.618 ± 0.210) than asymptomatic ones (0.227 ± 0.171). After excluding outliers, differences were more pronounced (H = 10.89; p < 0.01). Underdrainage patients had intermediate ΔP2/P1 values (0.387 ± 0.179) and consistently higher P2/P1 averages (>1.3). ROC analysis indicated that ΔP2/P1 > 0.3 suggested shunt dysfunction (AUC = 0.731), while the highest P2/P1 offered stronger discrimination (AUC = 0.782). A global average P2/P1 > 1.3 was linked to underdrainage, with the lowest P2/P1 values differentiating overdrainage (0.948 ± 0.321) from underdrainage (1.143 ± 0.156). Conclusions: nICPw monitoring with the B4C system demonstrated potential for detecting shunt dysfunction. Combining parameters, especially ΔP2/P1 and highest P2/P1, improves diagnostic accuracy, offering a non-invasive method that may aid in distinguishing normal from abnormal shunt function.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Shunt Dysfunction Assessment in Shunted Patients via Multiparametric Non-Invasive ICP Waveform Monitoring
Date Crossref
07/10/2025
Éditeur
MDPI AG
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Traumatic Brain Injury and Neurovascular DisturbancesCerebrospinal fluid and hydrocephalusCardiovascular Health and Disease Prevention

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.