Bench evaluation of six non-invasive ventilation home ventilators: comparison with an ICU ventilator and unsupervised clustering

5Citations signalées — pas une note de qualité
6Institutions déclarées
2Pays d’affiliation déclarés

Résumé fourni par la source

Introduction Chronic hypercapnic respiratory failure often necessitates non-invasive ventilation (NIV) at home. Our study aimed to assess the static and dynamic performance of six modern NIV home ventilators and one intensive care unit (ICU) ventilator and to identify performance clusters among the devices. Methods A two-compartment lung model was connected to seven NIV ventilators (Sefam Stan, Philips A40, Philips DreamStation, Resmed Lumis 150, Löwenstein PrismaVent 30C, Löwenstein PrismaVent 40 and BellaVista 1000) in pressure-support mode. Static and dynamic (triggering and pressurisation) performances were assessed through three distinct clinical phenotypes and four levels of unintentional leak. Clustering analysis was performed using K-means. Results For each of the seven ventilators, 144 conditions were tested, and a total of 3024 cycles were analysed. Static and dynamic performances were good to excellent across home ventilators, significantly higher than the ICU ventilator. Clustering analysis identified three performance clusters. Cluster 1 (Sefam Stan and Philips A40) showed significantly more precise accuracy of inspiratory pressure than Cluster 2 (Philips DreamStation, Resmed Lumis 150, Löwenstein PrismaVent 30C and Löwenstein PrismaVent 40) and Cluster 3 (BellaVista 1000): mean error=4.3%±5.1% versus 8.5%±6.7% versus 10.6%±14.7% respectively, p<0.001. For the triggering delay, Cluster 1 displayed shorter delays than Cluster 2 and Cluster 3 (41±5 ms vs 58±11 ms vs 67±13 ms, respectively, p<0.001). For the pressurisation delay, Cluster 1 displayed shorter delays than Cluster 2 and Cluster 3 (42±6 ms vs 64±14 ms vs 87±14 ms, respectively, p<0.001). For the pressure-time product at 300 ms, Cluster 1 displayed higher area under the curve for the first 300 ms than Cluster 2 and Cluster 3 (2.1±1.1 cmH 2 O /s vs 1.6±0.8 cmH 2 O/s vs 1.3±1.0 cmH 2 O/s, respectively, p<0.001). Continuous unintentional leaks did not modify the pressurisation performances in Cluster 1 but altered them in Clusters 2 and 3. Conclusion The six NIV home ventilators demonstrated superior performance compared with the tested ICU ventilator. The ventilators of Cluster 1 were identified as top performers in clustering analysis and compensated for unintentional continuous leaks.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Bench evaluation of six non-invasive ventilation home ventilators: comparison with an ICU ventilator and unsupervised clustering
Date Crossref
01/10/2025
Éditeur
BMJ
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Respiratory Support and MechanismsNosocomial Infections in ICUMechanical Circulatory Support Devices

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.