Cyclosporin A Monitoring in Allogeneic Hematopoietic Stem-Cell Transplant Recipients: A Comparison Between Siemens Viva-ProE System and LC-MS/MS

0Citations signalées, ce qui n’est pas une note de qualité
5Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : cn. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: Allogeneic hematopoietic stem-cell transplantation is a critical treatment for blood-related diseases often complicated by graft-versus-host disease and managed using cyclosporin A (CsA). Therapeutic drug monitoring is essential due to the narrow therapeutic window of CsA. METHODS: In this study, a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the quantification of whole-blood CsA concentrations was developed and validated. CsA concentration in allogeneic hematopoietic stem-cell transplantation patients was measured using the newly established LC-MS/MS and the enzyme multiplied immunoassay technique-based Siemens viva-ProE system. The results from both methods were statistically analyzed across different concentration ranges to evaluate method agreement, correlation, and clinical implications. RESULTS: The established LC-MS/MS method demonstrated excellent linearity (r 2 = 0.9983) within the CsA concentration range of 10-1000 ng/mL, with the validation parameters meeting the bioanalytical method validation criteria. Analysis of 161 samples revealed a strong correlation (r = 0.9908) between the methods; however, Siemens viva-ProE system overestimated CsA levels by an average of 8.7%, with a greater bias at higher concentrations. Correction through Deming regression significantly improved the agreement at therapeutically relevant thresholds. CONCLUSIONS: The developed LC-MS/MS assay demonstrated acceptable performance in measuring CsA concentration in whole blood. The comparison between the 2 methods highlights the need for method specification in therapeutic drug monitoring and provides regression equations to enhance the monitoring accuracy in clinical practice.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Cyclosporin A Monitoring in Allogeneic Hematopoietic Stem-Cell Transplant Recipients: A Comparison Between Siemens Viva-ProE System and LC-MS/MS
Date Crossref
29/09/2025
Éditeur
Ovid Technologies (Wolters Kluwer Health)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Hematopoietic Stem Cell TransplantationRenal Transplantation Outcomes and TreatmentsAcute Myeloid Leukemia Research

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.