Cardiac DTI Inline Automated Processing and On-scanner Diffusion Parameter Map Visualization Using FIRE
Le résumé fourni par la source
Motivation: Cardiac DTI provides a non-invasive tool to assess microstructure in the heart; however, its clinical use remains hindered by offline post-processing. Goal(s): To achieve inline automated processing and on-scanner results visualization for cardiac DTI. Approach: Inline automated post-processing pipeline was implemented using FIRE incorporating motion correction, diffusion tensor fitting, and automatic myocardial center detection for helix angle map generation. Results: Inline processing time on scanner for 5 slices was 1 minute. Diffusion parameter maps were successfully generated prospectively in 28 volunteers and 19 patients, except for myocardial center detection failure in 2 slices from one volunteer. Impact: Inline processing enables evaluation of diffusion parameters immediately on scanner or in PACS and enables wide-spread distribution and clinical evaluation of cardiac DTI. Thus, this work implements a key step towards clinical use of cardiac DTI.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Cardiac DTI Inline Automated Processing and On-scanner Diffusion Parameter Map Visualization Using FIRE
- Date Crossref
- 16/09/2025
- Éditeur
- ISMRM
- Type
- proceedings-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.