Comparison of consistency between wireless and conventional wired monitoring systems in gastrointestinal endoscopy

1Citations signalées, ce qui n’est pas une note de qualité
3Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : cn. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

There is a lack of research on use and effectiveness of wearable monitoring devices in the setting of invasive procedures, such as gastrointestinal endoscopy. Our study aimed to compare the consistency between wireless and conventional wired monitoring systems in gastrointestinal endoscopy under sedation. One hundred and twenty patients undergoing gastrointestinal endoscopy with sedation/anesthesia were fitted with both wireless and conventional wired monitoring systems. Multiple physiological parameters, including heart rate (HR), pulse rate, peripheral oxygen saturation (SpO2) and systolic (SBP), diastolic (DBP) and mean blood pressure (MAP), were monitored and recorded simultaneously. The correlation coefficients were 0.999, 0.939, 0.606, 0.837, 0.934 and 0.941 for HR, pulse rate, SpO2, SBP, DBP and MAP, respectively (all p < 0.001). The HR exceeded its limits of agreements (LoAs) and clinically acceptable consistency thresholds in 56 (4.17%) and 41 (3.05%) data, pulse rate in 84 (6.0%) and 137 (9.80%) data, SpO2 in 116 (8.30%) and 18 (1.28%) data, SBP in 39 (6.32%) and 121 (19.61%)data, DBP in 34 (5.34%) and 52 (8.43%) data, and MAP in 28 (4.54%) and 57 (9.24%) data. The polar concordance rate at 30 degrees was 59.31% for SBP, 61.12% for DBP, and 68.55% for MAP according to polar plot trending analysis. This study demonstrated that the wireless monitoring system showed adequate agreement with wired system for HR and SpO₂, but suboptimal agreement for pulse rate, SBP, DBP, and MAP during gastrointestinal endoscopy. The blood pressure measurements meets the ISO 81060-2 accuracy criteria. Overall, wireless monitoring is generally acceptable for gastrointestinal endoscopy, but further validation is needed for high-risk patients.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Comparison of consistency between wireless and conventional wired monitoring systems in gastrointestinal endoscopy
Date Crossref
03/08/2025
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Surgical Simulation and Training

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.