Comparative evaluation of a novel card-based spacer (MDI PLUS®) versus traditional aerochamber plus®: In-vivo and ex-vivo bioavailability study

1Citations signalées — pas une note de qualité
4Institutions déclarées
3Pays d’affiliation déclarés

Résumé fourni par la source

BACKGROUND: Effective aerosol delivery via pressurized metered dose inhalers (pMDIs) is often compromised by poor actuation-inhalation coordination. Spacer devices are commonly used to improve lung deposition. This study aimed to compare the lung and systemic bioavailability of salbutamol delivered via pMDI alone, with a traditional AeroChamber Plus® spacer, and with a novel disposable MDI Plus® carton spacer. METHODS: In a randomized, three-period crossover study, 20 healthy adults received 5 puffs (100 µg/puff) of salbutamol via each inhalation setup. Urine samples were collected 30 min post-inhalation (for lung bioavailability) and over 24 h (for systemic absorption). An ex-vivo setup using inhalation filters was also employed to quantify total delivered dose. RESULTS: The ex-vivo mean ± SD total collected dose of salbutamol was significantly higher with pMDI alone (352.35±23.41 µg) than AeroChamber Plus (196.57±13.32 µg, p < 0.001) and MDI Plus (182.86±11.34 µg, p < 0.001). Urinary salbutamol values at 30 min were significantly greater with AeroChamber Plus (14.61±4.36 µg, p < 0.001) and MDI Plus (10.97±3.83 µg, p < 0.01) compared to pMDI alone (5.92±1.07 µg). Urinary salbutamol at 24 h showed higher systemic bioavailability with pMDI alone (139.09±15.68 µg) versus when connected with both AeroChamber Plus and MDI Plus, 79.57±19.48, and 59.26±18.12 µg, p < 0.001, respectively. CONCLUSION: Compared to the pMDI alone, both spacers significantly enhanced lung delivery and reduced systemic absorption. Whilst the AeroChamber Plus® delivered slightly more medication to the lung, the MDI Plus® spacer offers a cost-effective alternative that is suitable for emergencies and outdoor use.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Comparative evaluation of a novel card-based spacer (MDI PLUS®) versus traditional aerochamber plus®: In-vivo and ex-vivo bioavailability study
Date Crossref
01/09/2025
Éditeur
Elsevier BV
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Inhalation and Respiratory Drug DeliveryAsthma and respiratory diseasesNeonatal Respiratory Health Research

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.