Aller au contenu principal
Accès ouvert déclaré 2025 article

Estimates of cardiac implanted electronic device neutron dose for a pencil beam scanning proton therapy system

0Citations signalées, ce qui n’est pas une note de qualité
5Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: The presence of a cardiovascular implantable electronic device (CIED) is frequently viewed as a contraindication to proton therapy due to the creation of secondary neutrons that can potentially damage CIED electronics. As a result, photon therapy is typically recommended for patients with CIEDs. PURPOSE: The study aims to provide a method for estimating neutron dose to a CIED by measuring equivalent neutron dose at varying distances from isocenter and field edge. This estimation can be used to guide clinical decisions by balancing the risk of neutron-induced CIED damage against the therapeutic benefits of proton therapy. METHODS: Standardized three-dimensional measurement fields of varying dimensions were delivered using the IBA ProteusONE (Ion Beam Applications SA, Walloon Brabant, Belgium) pencil beam scanning proton therapy system, with each field delivering an RBE-weighted dose of 2 Gy. Baseline measurement included one treatment field being delivered with a range shifter. BD-PND detectors (Bubble Technology Industries, Chalk River, ON) were placed at a defined distances from the surface mark of beam isocenter and field edge to record neutron doses. Additionally, clinical treatment plans including two prostate/pelvis plans, two head & neck plans, two brain plans, and one breast plans were delivered, with detectors placed at positions corresponding to the location of a CIED. RESULTS: field increased the dose to 0.66 µSv/2 Gy at 50 cm to 14.2 µSv/2 Gy at 10 cm from isocenter. For clinical treatment plans, neutron doses ranging from 0.14 µSv/2 Gy to 9.5 µSv /2 Gy at 24-56 cm from isocenter. CONCLUSION: These measurements provide a foundation for estimating neutron doses to CIEDs, enabling physicists and physicians to evaluate the feasibility of proton therapy for patients with CIEDs. The results support informed clinical decision-making by quantifying the risk of neutron-induced damage relative to the therapeutic benefits of proton therapy.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Estimates of cardiac implanted electronic device neutron dose for a pencil beam scanning proton therapy system
Date Crossref
01/07/2025
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Radiation Therapy and DosimetryAdvanced Radiotherapy TechniquesBoron Compounds in Chemistry

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.