Hybrid proton planning combining spread-out Bragg peak beams with transmission beams to shorten field delivery times while maintaining plan quality
Rattachement africain : us, be, nl. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Background and purpose: Delivery times of Bragg-peak-based intensity-modulated proton therapy fields play an important role in patient throughput and comfort. Despite the associated exit dose, single-layer proton transmission beams benefit from sharper penumbras and are extremely fast to deliver. In this study, we investigated the trade-offs in field delivery times (FDT) and plan quality when using both field types. Materials and methods: Reference treatment plans were created for eight left-sided breast and ten oropharynx cancer patients using an in-house automatic iterative optimizer. Comparative hybrid plans were created for each case. For breast, different combinations of transmission and conventional beams were evaluated. For oropharynx, two of the three conventional beams were replaced with six transmission beams. Hybrid plans were evaluated by comparing the dose metrics and FDT against the reference plans. Results: Hybrid breast plans exhibited mean and maximum organ at risk (OAR) doses, and target dose homogeneity and conformity comparable to the reference plans, while their FDTs decreased by median (interquartile range) of 58 %/166 s (56-61 %). Compared to reference plans, hybrid oropharynx plans exhibited higher mean OAR dose especially to oral cavity (median of 34 Gy vs 31 Gy) and spinal cord (20 Gy vs 11 Gy), while FDTs decreased by 73 %/91 s (71-73 %). Discussion: Depending on the case, hybrid planning can significantly reduce total FDT with only limited impact on plan quality. The reduced total FDT can improve patient comfort, reduce overall duration of the treatment, and improve beam scheduling at multi-room centers.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Hybrid proton planning combining spread-out Bragg peak beams with transmission beams to shorten field delivery times while maintaining plan quality
- Date Crossref
- 01/07/2025
- Éditeur
- Elsevier BV
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Où se fait cette recherche
-
Siemens Healthcare (United States) pays non établi dans la noticeEntreprise
-
Siemens (Belgium) pays non établi dans la noticeEntreprise
-
Amsterdam University Medical Centers pays non établi dans la noticeÉtablissement de santé
-
a Siemens Healthineers Company Varian pays non établi dans la noticeEntreprise
-
Amsterdam UMC pays non établi dans la noticeInstitution
Siemens Healthcare (United States), Siemens (Belgium) et Amsterdam University Medical Centers, avec 2 autres affiliations.
Une affiliation ne permet pas de déduire la nationalité d’un auteur.