Multi‐Institutional Survey of 4‐Year Ultrasound Quality Control Findings

1Citations signalées, ce qui n’est pas une note de qualité
11Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

OBJECTIVES: Periodic quality control (QC) testing of ultrasound (US) imaging systems is essential to ensure and maintain image quality and safety. The study aims to analyze QC findings from medical physics annual surveys of modern clinical US systems in a multi-institutional survey. METHODS: QC results from annual surveys between 2018 and 2021 were retrospectively collected from 12 medical physicists from 11 institutions or consulting companies (hereafter referred to as sites). QC tests were classified as scanner- and transducer-related tests, where scanner-related tests also included evaluation of diagnostic workstation display monitors and routine QC programs if applicable. Test methodology and pass/fail criteria were established by each site. QC findings were defined as results requiring follow-up action by service engineers, vendors, or other personnel. The percentage of annual findings was calculated for each site and averaged across the years of data contributed. Furthermore, findings across all sites were aggregated for an overall analysis. RESULTS: QC data from a total of 1069 scanner tests and 4542 transducer tests were collected. The average annual percentage of scanner-related findings varied from 0 to 40.8% among all sites (first quartile 8.7%, median 18.2%, and third quartile 34.1%), while that of transducer-related findings ranged between 0.5% and 19.9% (first quartile 3.9%, median 7.2%, and third quartile 14.5%). For scanner-related tests, the top 3 categories of findings were associated with physical and mechanical integrity (171 findings, 74.7%), quantitative testing of the scanner display monitor (40 findings, 17.5%), and scanner port creating artifacts (10 findings, 4.4%). The top 3 findings for transducers related to uniformity and artifact (227 findings, 79.9%), physical and mechanical integrity (44 findings, 15.5%), and sensitivity in fundamental mode (9 findings, 3.2%). CONCLUSIONS: Medical physics annual surveys revealed considerable actionable findings of modern US systems, which inform the status of current US QC practices in the United States and may guide future standardization and recommendation of QC tests.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Multi‐Institutional Survey of 4‐Year Ultrasound Quality Control Findings
Date Crossref
03/07/2025
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Ultrasound Imaging and ElastographyMedical Device Sterilization and DisinfectionUltrasound and Hyperthermia Applications

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.